Published August 2, 2026 | Version v1
Dataset

Flow Battery Reproducibility Study – Participant 04 (Phase 2b)

  • 1. STFC

Description

Submission from participant 04 as part of the Flow Battery Reproducibility Study, Phase 2b. Job role: Postgrad/PhD student. Experience: < 6 months.

Additional details

Domain Specific Metadata

 
Property Value
Fb Data Collection Phase Phase 2b
Fb Job Role Postgrad/PhD student
Fb Experience Flow Batteries < 6 months
Fb Participant Identifier 04
Fb Number Of Repeats 3
Fb Electrode Pretreatment Type Heat-treated
Fb Heat Treatment Furnace Type Muffle
Fb Heat Treatment Atmosphere Air
Fb Heat Treatment Ramp Rate 10.0
Fb Heat Treatment Dwell Temperature 400.0
Fb Heat Treatment Dwell Time 24.0
Fb Heat Treatment Cooling Method Passive
Fb Heat Treatment Cooling Details None
Fb Heat Treatment Batching Treated individually before each test
Fb Time Between Heat Treatment And Use 24.0
Fb Electrode Storage After Heat Treatment Enclosed (non-inert)
Fb Electrode Storage After Receipt Enclosed (non-inert)
Fb Membrane Storage Type Wet
Fb Membrane Storage Liquid deionised H2O
Fb Membrane Storage Details Stored in closed beaker
Fb Membrane Pretreatment Yes
Fb Ferrocyanide Chemical Details Potassium hexacyanoferrate(II) trihydrate, K4Fe(CN)6 · 3H2O
Fb Ferrocyanide Purity 99.0
Fb Ferrocyanide Supplier Sigma Aldrich
Fb Ferricyanide Chemical Identity Potassium ferricyanide(III), K3Fe(CN)6
Fb Ferricyanide Purity 99.0
Fb Ferricyanide Supplier Sigma Aldrich
Fb Supporting Electrolyte Identity Potassium chloride, KCl
Fb Potassium Chloride Purity 99.0
Fb Potassium Chloride Supplier Sigma Aldrich
Fb Water Type Deionised
Fb Water Purification System Milli-Q (Merck)
Fb Water Purification Details EQ 7008
Fb Water Resistivity 18.2
Fb Electrolyte Preparation Strategy Multiple stock batches used across repeats
Fb Electrolyte Batch Size 500.0
Fb Electrolyte Storage Environment Dark (cupboard/covered)
Fb Electrolyte Storage Duration 730.0
Fb Electrolyte Volume Measurement Method Volumetric
Fb Reagent Purity Used No
Fb Water Ph Measured No
Fb Electrolyte Ph Measured No
Fb Electrode Cutting Method Template-guided manual cutting
Fb Potentiostat Usage Different instruments for some/all tests
Fb Potentiostat Manufacturer Cycling Ivium | other (please specify)
Fb Potentiostat Model Cycling Octostat5000
Fb Potentiostat Manufacturer Polarisation Ivium
Fb Potentiostat Model Polarisation Octostat5000
Fb Potentiostat Manufacturer Impedance Metrohm Autolab
Fb Potentiostat Model Impedance PGSTAT302N
Fb Potentiostat Calibration Frequency Yearly
Fb Potentiostat Calibration Method Dummy cell (manufacturer supplied)
Fb Connection Method Crocodile/alligator clips
Fb Electrical Connection Scheme 4-point (4-wire)
Fb Potentiostat Cable Length 1–2 m
Fb Potentiostat Cable Type Manufacturer cable – standard length
Fb Potentiostat Cable Shielding Mixed shielding
Fb Number Of Pumps Single (including dual head)
Fb Pump Type Peristaltic
Fb Pump Manufacturer Ismatec
Fb Pump Model Reglo ICC
Fb Reservoir Material Glass (borosilicate)
Fb Reservoir Type Duran/Schott bottle (GL bottle)
Fb Reservoir Volume 110.0
Fb Electrolyte Volume Used 100
Fb Impedance Reservoir Config Single reservoir
Fb Polarisation Reservoir Config Single reservoir
Fb Temperature Control Ambient (not controlled)
Fb Laboratory Temperature 22–24 °C
Fb Pump Flow Rate Calibrated Yes
Fb Calibration With Cell Yes
Fb Calibration Fluid Electrolye
Fb Pump Recalibrated No
Fb Pump Calibration Method Manual gravimetric
Fb Leak Test Performed Yes
Fb Break In Performed Yes
Fb Break In Duration 1.0
Fb Break In Flow Rate 25
Fb Cell Assembly Issues The membrane gaskets started to poke out of the side of the cell on the later runs leading to some leaks (I think this was the source)
Fb Additional Experiments Yes
Fb Repeat Method Full cell disassembly and rebuild, fresh electrolyte
Fb Electrolyte Leakage Reported No
Fb Typical Group Practice Leaks Discard data if any leak occurs | Case-by-case decision | other (please specify)
Fb Electrolyte Stirring Yes
Fb Reservoir Stirring Method Magnetic stir bar
Fb Stirring Rate Not recorded
Fb Electrolyte Sparging Yes
Fb Sparging Gas Nitrogen
Fb Gas Humidification Humidified
Fb Sparging Application Continuous sparging
Fb Sparging Gas Flow Rate Not recorded
Fb Polarisation Beyond Limits No
Fb Impedance Control Mode Potentiostatic (PEIS)
Fb Consent Public Display Yes